B.2 Normal test environment
38.176-13GPPIntegrated Access and Backhaul (IAB) conformance testingNRPart 1: conducted conformance testingRelease 17TS
When a normal test environment is specified for a test, the test should be performed within the minimum and maximum limits of the conditions stated in table B.1.
Table B.1: Limits of conditions for normal test environment
Condition |
Minimum |
Maximum |
Barometric pressure |
86 kPa |
106 kPa |
Temperature |
15 °C |
30 °C |
Relative humidity |
20 % |
85 % |
Power supply |
Nominal, as declared by the manufacturer |
|
Vibration |
Negligible |
The ranges of barometric pressure, temperature and humidity represent the maximum variation expected in the uncontrolled environment of a test laboratory. If it is not possible to maintain these parameters within the specified limits, the actual values shall be recorded in the test report.
NOTE: This may, for instance, be the case for measurements of radiated emissions performed on an open field test site.